AMETEK Lx Series II Programming Manual User Manual

Page 88

background image

Programming Manual

Lx \ Ls Series II

84

Query Syntax

[SOURce:]LIST:TTLTrg:POINts?

Returned Parameters

Example

LIST:TTLT:POIN?

Related Commands

LIST:TTLT

LIST:VOLTage

This command specifies the output voltage points in a list. The voltage points are given in the
command parameters, which are separated by commas. The order in which the points are
entered determines the sequence in which the list will be output when a list transient is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.

The maximum peak voltage that the AC source can output is 425 V peak. This includes any
combination of voltage, voltage offset, and function shape values. Therefore, the maximum value
that can be programmed depends on the peak-to-rms ratio of the selected waveform. For a
sinewave, the maximum voltage that can be programmed is 300 V rms.

Note: You cannot program a voltage that produces a higher volt-second on the output than a

300V rms sinewave.

Command Syntax

[SOURce:]LIST:VOLTage[:LEVel] {,}

Parameters

0 to 300 (for sinewaves)

Unit

V (rms voltage)

Examples

LIST:VOLT 2.0,2.5,3.0

LIST:VOLT MAX,2.5,MIN

Query Syntax

[SOURce:]LIST:VOLTage[:LEVel]?

Returned Parameters

Related Commands

LIST:VOLT:POIN? LIST:COUN LIST:DWEL LIST:STEP

LIST:SHAP LIST:VOLT:OFFS

LIST:VOLTage:POINts?

This query returns the number of points specified in LIST:VOLT. Note that it returns only the total
number of points, not the point values.

Query Syntax

[SOURce:]LIST:VOLTage:POINts?

Returned Parameters

Example

LIST:VOLT:POIN?

Related Commands

LIST:VOLT

LIST:VOLTage:SLEW

This command specifies the output offset slew points in a list. The slew points are given in the
command parameters, which are separated by commas. The order in which the points are
entered determines the sequence in which the list will be output when a list transient is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.

Command Syntax

[SOURce:]LIST:VOLTage:SLEW {,}

Parameters

1E-4 to 9.9E37 | INFinity | MINimum | MAXimum

Unit

V/S (volts per second)

Example

LIST:VOLT:SLEW 10, 1E2, INF

Query Syntax

[SOURce:]LIST:VOLTage:SLEW?

Returned Parameters

Related Commands

LIST:VOLT:SLEW:POIN? LIST:COUN LIST:DWEL LIST:STEP

LIST:VOLTage:SLEW:POINts?

This query returns the number of points specified in LIST:VOLTage:SLEW. Note that it returns
only the total number of points, not the point values.