AMETEK Lx Series II Programming Manual User Manual
Page 88
![background image](https://www.manualsdir.com/files/771680/content/doc088.png)
Programming Manual
Lx \ Ls Series II
84
Query Syntax
[SOURce:]LIST:TTLTrg:POINts?
Returned Parameters
Example
LIST:TTLT:POIN?
Related Commands
LIST:TTLT
LIST:VOLTage
This command specifies the output voltage points in a list. The voltage points are given in the
command parameters, which are separated by commas. The order in which the points are
entered determines the sequence in which the list will be output when a list transient is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.
The maximum peak voltage that the AC source can output is 425 V peak. This includes any
combination of voltage, voltage offset, and function shape values. Therefore, the maximum value
that can be programmed depends on the peak-to-rms ratio of the selected waveform. For a
sinewave, the maximum voltage that can be programmed is 300 V rms.
Note: You cannot program a voltage that produces a higher volt-second on the output than a
300V rms sinewave.
Command Syntax
[SOURce:]LIST:VOLTage[:LEVel]
Parameters
0 to 300 (for sinewaves)
Unit
V (rms voltage)
Examples
LIST:VOLT 2.0,2.5,3.0
LIST:VOLT MAX,2.5,MIN
Query Syntax
[SOURce:]LIST:VOLTage[:LEVel]?
Returned Parameters
Related Commands
LIST:VOLT:POIN? LIST:COUN LIST:DWEL LIST:STEP
LIST:SHAP LIST:VOLT:OFFS
LIST:VOLTage:POINts?
This query returns the number of points specified in LIST:VOLT. Note that it returns only the total
number of points, not the point values.
Query Syntax
[SOURce:]LIST:VOLTage:POINts?
Returned Parameters
Example
LIST:VOLT:POIN?
Related Commands
LIST:VOLT
LIST:VOLTage:SLEW
This command specifies the output offset slew points in a list. The slew points are given in the
command parameters, which are separated by commas. The order in which the points are
entered determines the sequence in which the list will be output when a list transient is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.
Command Syntax
[SOURce:]LIST:VOLTage:SLEW
Parameters
1E-4 to 9.9E37 | INFinity | MINimum | MAXimum
Unit
V/S (volts per second)
Example
LIST:VOLT:SLEW 10, 1E2, INF
Query Syntax
[SOURce:]LIST:VOLTage:SLEW?
Returned Parameters
Related Commands
LIST:VOLT:SLEW:POIN? LIST:COUN LIST:DWEL LIST:STEP
LIST:VOLTage:SLEW:POINts?
This query returns the number of points specified in LIST:VOLTage:SLEW. Note that it returns
only the total number of points, not the point values.