Command Syntax
[SOURce:]LIST:SHAPe
Parameters
SINusoid|SQUare|CSINusoid|
Examples
LIST:SHAP
Query Syntax
[SOURce:]LIST:SHAP?
Returned Parameters
Related Commands
LIST:SHAP:POIN? LIST:COUN LIST:DWEL LIST:STEP LIST:VOLT
LIST:SHAPe:POINts?
This query returns the number of points specified in LIST:SHAP. Note that it returns only the total
number of points, not the point values.
Query Syntax
[SOURce:]LIST:SHAPe:POINts?
Returned Parameters
Example
LIST:SHAP:POIN?
Related Commands
LIST:SHAP
LIST:STEP
This command specifies how the list sequencing responds to triggers.
ONCE causes the list to advance only one point after each trigger. Triggers that arrive during a
dwell delay are ignored.
AUTO causes the entire list to be output sequentially after the starting trigger, paced by its dwell
delays. As each dwell delay elapses, the next point is immediately output.
Command Syntax
[SOURce:]LIST:STEP
Parameters
ONCE | AUTO
*RST Value
AUTO
Examples
LIST:STEP ONCE
Query Syntax
[SOURce:]LIST:STEP?
Returned Parameters
Related Commands
LIST:COUN
LIST:DWEL
LIST:TTLTrg
This command sets the sequence of Trigger Out list points. Each point which is set ON will cause
a pulse to be output at Trigger Out when that list step is reached. Those entries which are set
OFF will not generate Trigger Out pulses. The order in which the list points are given determines
the sequence in which Trigger Out pulses will be output when a list transient is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.
Command Syntax
[SOURce:]LIST:TTLTrg
Parameters
0 | 1 | OFF | ON
Examples
LIST:TTLT 1,0,1 LIST:TTLT ON,OFF,ON
Query Syntax
LIST:TTLT?
Returned Parameters
0 | 1
Related Commands
LIST:TTLT:POIN? LIST:COUN LIST:DWEL LIST:STEP
OUTP:TTLT:STAT OUTP:TTLT:SOUR
LIST:TTLTrg:POINts?
This query returns the number of points specified in LIST:TTLT. Note that it returns only the total
number of points, not the point values.