TRIGger

When the trigger subsystem has been initiated, the TRIGger command generates a trigger signal
regardless of the selected trigger source.

Command Syntax

TRIGger[:SEQuence1][:IMMediate]

TRIGger[:TRANsient][:IMMediate]

Parameters

None

Examples

TRIG TRIG:TRAN TRIG:SEQ1:IMM

Related Commands

ABOR TRIG:SOUR TRIG:DEL TRIG:SYNC

TRIG:SYNC:PHAS INIT INIT:CONT *TRG *WAI

TRIGger:DELay

This command sets the time delay between the detection of a trigger signal and the start of any
corresponding trigger action. After the time delay has elapsed, the trigger is implemented unless
the trigger system is also waiting for a sync signal that has been specified by
TRIGger:SYNChronous:PHASe.

Command Syntax

TRIGger[:SEQuence1]:DELay

TRIGger[:TRANsient]:DELay

Parameters

3-phase models: 0 to 1.07533E6|MINimum|MAXimum

1-phase models: 0 to 4.30133E5|MINimum|MAXimum

Unit

s (seconds)

*RST Value

0

Examples

TRIG:DEL .25 TRIG:DEL MAX TRIG:TRAN:DEL 1

Query Syntax

TRIGger[:SEQuence1]:DELay?

TRIGger[:TRANsient]:DELay?

Returned Parameters

Related Commands

ABOR TRIG

TRIG:SOUR

TRIG:SYNC

TRIG:SYNC:PHAS

INIT

INIT:CONT

*TRG *WAI

TRIGger:SOURce

This command selects the trigger source for the first sequence in generating a step, pulse, or list
output as follows:

BUS

IEEE-488 device, *TRG, or (Group Execute Trigger)

EXTernal

AC source's backpanel Trigger In1 SMA connector

IMMediate

trigger is generated as soon as the trigger system is initiated.

Command Syntax

TRIGger[:SEQuence1]:SOURce

TRIGger[:TRANsient]:SOURce

Parameters

BUS | EXTernal | IMMediate

*RST Value

BUS

Examples

TRIG:SOUR BUS

TRIG:TRAN:SOUR EXT

Query Syntax

TRIGger[:SEQuence1]:SOURce?

TRIGger[:TRANsient]:SOURce?

Returned Parameters

Related Commands

ABOR TRIG

TRIG:DEL

TRIG:SYNC

TRIG:SYNC:PHAS

INIT

INIT:CONT *TRG *WAI