AMETEK Lx Series II Programming Manual User Manual
Page 65
![background image](https://www.manualsdir.com/files/771680/content/doc065.png)
Programming Manual
Lx \ Ls Series II
61
Query Syntax
MEASure[:SCALar]:VOLTage:HARMonic:THD?
FETCh[:SCALar]:VOLTage:HARMonic:THD?
Parameters
None
Examples
MEAS:VOLT:HARM:THD?
FETC:VOLT:HARM:THD?
Returned Parameters
Related Commands
INST:NSEL
MEASure:THDistortion:MODE
This command sets the calculation method for THD measurements. The distortion calculation is
based on the H2 through H50 with the RMS voltage or current in the denominator. Note that some
definitions of THD use the fundamental component (H1) of the voltage or as the denominator.
Lx/Ls units with firmware revision 0.88 or higher can be programmed to use the fundamental
component as the denominator instead of the RMS value. This mode can only be programmed
over the bus by sending the “MEAS:THD:MODE FUND” command. At power up or after a reset
command, the mode will revert back to the RMS mode. This mode setting is not saved in any of
the set up registers.
Syntax
MEASure:THDistortion:MODE
Parameters
RMSQuare | FUNDamental
Examples
MEAS:THD:MODE FUND
Query Syntax
MEAS:THD:MODE?
Returned Parameters
Related Commands
MEAS:VOLT:HARM:THD?
MEAS:CURR:HARM:THD?