6 meas ipol - fixed endpoint evaluation – Metrohm tiamo 2.3 Manual User Manual
Page 986
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
970
■■■■■■■■
tiamo 2.3
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-
[New]
Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 928).
[Properties]
Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 928).
[Delete]
Delete the external measured value selected in the table.
5.6.4.5.6
MEAS Ipol - Fixed endpoint evaluation #
Dialog window: Method
▶ MEAS Ipol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation #
Measured quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value (Titrando, 867, 855)
Range
-2000.0 ... 2000.0 mV
Measured value (Titrino)
Range
-2000 ... 2000 mV
Time
Range
0.0 ... 999999.9 s