5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1130
■■■■■■■■
tiamo 2.3
Temperature
The section Temperature is only displayed for Device type = 089 Pho-
tometer.
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to continuous, then the temper-
ature will be measured continuously.
Range
20.0 ... 60 °C
Default value
25.0 °C
5.6.5.12
CAL LOOP pH
5.6.5.12.1
CAL LOOP pH - Overview
Dialog window: Method
▶ CAL LOOP pH ▶ Properties... ▶ CAL LOOP pH - 'Com-
mand name'
Loop command for the calibration of pH electrodes which is used
together with the command CAL MEAS pH .
Appearance
The command has the following appearance:
Parameters
The parameters for the command CAL LOOP pH are configured in the
following dialog window:
■
Command variables
The following command variables are generated in the method run by the
command CAL LOOP pH and can be used in formulas under the desig-
nation 'Command name.Variable designation':
Desig-
nation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.ENP
Electrode zero point calculated from the calibration