8 stat ph - evaluations, Evaluations – Metrohm tiamo 2.3 Manual User Manual
Page 897
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
881
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Track call on limit exceeding
In this table, which cannot be edited directly, a maximum of 20 entries
can be defined as to which track is to be started automatically when a
particular limit value is exceeded.
[New]
Open the dialog window Track call # in which the parameters for the
track call can be entered.
[Properties]
Open the dialog window Track call # in which the parameters for the
track call can be edited (see Chapter 5.6.3.8.4, page 857).
[Delete]
Delete the track call selected in the table.
5.6.3.8.6.8
STAT pH - Evaluations
Tab: Method
▶ STAT pH ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following methods for the evaluation of titration curves can be activa-
ted and defined on this tab:
Rate evaluation
Rate evaluation
on | off (Default value: off)
If this check box is activated, than a maximum of 9 time slots can be
defined in which the mean dosing rate can be determined by linear
regression. The defined time slots are shown in the window table and can
be edited with the following buttons:
[New]
Open the dialog window Evaluation window # in which the parameters
for a new time slot can be entered (see Chapter 5.6.3.8.3, page 857).