5 Method
tiamo 2.3
■■■■■■■■
791
User
Editing the individual titration parameters which affect the titration
rate.
Control
Note
Will be displayed only if the option User is selected in the selection list
Titration rate.
Control range
The control range defines the measured value range before the given end-
point. It has a decisive influence on the titration rate and thus also on the
accuracy. Within the control range, dosing is controlled by the Minimum
rate. The closer the endpoint, the slower the dosing until the Minimum
rate has been reached. The larger the control range, the slower the titra-
tion. Outside the control range, dosing is carried out continuously at
Maximum rate.
Titrando
Range
0.1 ... 2000.0 mV
Default value
300.0 (slow), 100.0 (optimal), 30.0 (fast) mV
Titrino
Range
1 ... 2000 mV
Default value
300 (slow), 100 (optimal), 30 (fast) mV
Max. rate
Rate at which dosing is carried out outside of the Control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Titrando
Range
0.01 ... 166.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min