10 det ipol - volume window – Metrohm tiamo 2.3 Manual User Manual
Page 550
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
534
■■■■■■■■
tiamo 2.3
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Titrando, 855
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.4.10
DET Ipol - Volume window
Dialog window: Method
▶ DET Ipol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with volume window (ml) ▶ [New]/[Properties] ▶ Volume win-
dow #
Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-