Metrohm tiamo 2.3 Manual User Manual
Page 463
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
447
Display message
on | off (Default value: on)
If this option is enabled, all active tracks are stopped and the message
defined in the text box will be displayed in the event that the limits are
exceeded.
Record message
on | off (Default value: on)
If this option is enabled, the message defined in the text box will be docu-
mented in the determination if the limits are exceeded.
Message by E-mail
on | off (Default value: off)
If this option is enabled, the message defined in the text box will be sent
to the address defined under
[E-mail]
Opens the window Send E-mail for defining the E-mail parameters (see
Chapter 5.6.2.2.6, page 448).
Acoustic signal
on | off (Default value: off)
If this option is enabled, an acoustic signal will be emitted in addition to
the message defined above when a limit is exceeded.
Action
Note
Is displayed only for method variables of the types Number or Date/
Time.
Action
on | off (Default value: off)
If this option is enabled, one of the two following actions will be carried
out automatically in the event that the result limits are exceeded.
Selection
Stop determination | Stop determination and
series
Default value
Stop determination