5 Method
tiamo 2.3
■■■■■■■■
527
Evaluation without window
EP criterion
Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points:
Titrino
Selection
all | greatest | last | off
Default value
all
Titrando, 855
Selection
all | greatest | last | ascending | descending |
off
Default value
all
all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.
Evaluation with measured value window (Upol)
EP criterion
Criterion for the recognition of equivalence points which is valid for all
windows. Equivalence points whose ERC is smaller than the set EP crite-
rion will not be recognized.