5 Method
tiamo 2.3
■■■■■■■■
1205
Note
For Dosing device and Solution exactly the same settings have to be
selected as for the dosing device reserved earlier by another command,
otherwise the dosing device is not being released.
Note
If the tandem dosing is used in a command (e.g. ADD), then both dos-
ing devices used have to be released each with a separate command
RLS DOS.
5.6.7
Automation commands
5.6.7.1
Automation commands - Overview
Menu item: Method
▶ Paste ▶ New command... ▶ Automation
Commands for the operation of Sample Processors.
The following eight automation commands can be selected:
■
MOVE
Moving to a rack position or an external position.
■
SWING
Swinging of the robotic arm (with Swing Head only).
■
LIFT
Moving to a lift position.
■
PUMP
Switching on/off the connected or built-in pumps.
■
STIR
Controlling a connected stirrer.
■
RACK
Initialization of the rack attached.
■
HEATER
Controlling the oven temperature of the 774 Oven Sample Processor.
■
FLOW
Regulation of the gas flow of the 774 Oven Sample Processor.
■
RLS DEV
Release of a device for using it in other parallel running methods.