Metrohm tiamo 2.3 Manual User Manual
Page 1205
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
1189
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Monitoring Temperature
Monitoring Temperature
on | off (Default value: off)
If this option is switched on then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the temperature. When the temperature falls below this
value, the action Lower temperature limit exceeded is triggered.
Titrando,855
Range
-20.0 ... 150.0 °C
Default value
-20.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok is triggered.
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
150.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
500.0 °C