3.6 Subwindow Run
■■■■■■■■■■■■■■■■■■■■■■
138
■■■■■■■■
tiamo 2.3
■
Check solutions
Check whether the solutions required in the instrument-specific com-
mands are present and available.
■
Check sensors
Check whether the sensors required in the instrument-specific com-
mands are present and available.
Note
When it is in READY status, the start test can also be triggered man-
ually with the menu item Tools
▶ Run test.
4 - Conditioning (only for methods with conditioning)
Conditioning is started in the case of methods which contain commands
with switched-on conditioning. The requirement for this is that the Auto-
matic conditioning option is switched on in the START command.
Conditioning is not yet part of determination and causes no counters to
be incremented (Sample number, Statistics).
■
Conditioning condition not fulfilled
After conditioning starts, the status switches to COND BUSY. The
determination can be started in this status with [Start], even though
the conditioning requirement is not fulfilled. Conditioning can also be
canceled with [Stop] or interrupted with [Hold] and resumed with
[Cont]. If an error occurs during conditioning, then neither the error
track nor the exit track will be run through.
■
Conditioning condition fulfilled
If the conditioning requirement has been fulfilled, then the status will
switch to COND READY. The determination can be started in this sta-
tus with [Start]. Conditioning can however also be canceled with
[Stop] or interrupted with [Hold] and resumed with [Cont]. As long
as the determination has not yet been started, the conditioning
requirement will continue to be monitored on a continuous basis.
5 - Method start
The following checks and actions are triggered with method start:
■
Check sample data
Check whether the sample data entered corresponds to the defined
method variables and is valid.
■
Assign determination ID
A unique determination ID is assigned for the determination.