5 Method
tiamo 2.3
■■■■■■■■
733
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.4.4.10, page 735).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.4.4.10, page 735).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL