3 dos u - dosing parameters, Dosing parameters – Metrohm tiamo 2.3 Manual User Manual
Page 1200
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1184
■■■■■■■■
tiamo 2.3
5.6.6.3.5.3
DOS U - Dosing parameters
Tab: Method
▶ DOS U ▶ Properties... ▶ Dosing parameters
Command name
Name of the command.
Input
25 characters
The volume to be added, the dosing rate and the dosing time are defined
on this tab.
Dosing criterion
Selects the two parameters that can be preset for dosing. The third para-
meter is calculated automatically from the formula Volume = dosing
time * dosing rate and is not shown (see Chapter 5.6.6.3.2, page
1157).
Selection
Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time
Default value
Volume/dosing rate
Volume
Fixed volume that is to be dosed. This parameter is only displayed for
Dosing criterion = Volume/Dosing rate or Volume/Dosing time.
Range
0.00000 ... 99999.9 mL
Default value
10.0000 mL
Dosing rate
Rate at which dosing should take place. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing device
used. If the entered dosing rate is too high for the selected dosing device,
it will automatically be reduced during dosing to the largest possible
value. This parameter is only displayed for Dosing criterion = Volume/
Dosing rate or Dosing rate/Dosing time.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min