Sensor
Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.
Selection
Sensor name | pH electrode | Metal electrode |
ISE electrode
Default value
Metal electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8