5 brc - titration parameters, Titration parameters – Metrohm tiamo 2.3 Manual User Manual
Page 865
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
849
Range
1.0 ... 999.0 µg/min
Default value
15.0 (slow), 25.0 (optimal), 50.0 (fast) µg/min
Selection
minimum
Stop criterion
Stop criterion
Switches off titration when the endpoint has been reached and the stop
criterion has been met.
Selection
Drift | rel. drift | Time | drift & time | off
Default value
drift & time
Stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift or drift
& time.
Range
1 ... 999 µg/min
Default value
15 µg/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Range
0 ... 999 µg/min
Default value
5 µg/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time or drift & time.
Range
0 ... 999 s
Default value
40 s
5.6.3.7.5
BRC - Titration parameters
Tab: Method
▶ BRC ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.