7 Instruments
tiamo 2.3
■■■■■■■■
1531
Number of positions
Shows the number of positions on the rack. If no rack is in position then
the display will be empty.
Shift rate
Sample rack speed of rotation for manual operation.
Range
3 ... 20 °/s
Default value
20 °/s
[Rack Data]
Opens the dialog window Rack data (see Chapter 6.12.3.1, page 1451)
or Rack data (774) (see Chapter 6.12.3.6, page 1456) in which the data
of the rack attached can be displayed and edited.
[Initialize rack]
Initializes the attached rack. This resets the rack, the lift and the robotic
arm, reads out the rack code and transfers the corresponding rack data to
the Sample Processor.
7.12.6
Properties - MSB #
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ MSB #
On the tabs MSB # the properties of the connector and the devices con-
nected to are displayed.
Request for dosing device preparation
Selection when the request for carrying out the command PREP (see
Chapter 5.6.6.5.1, page 1197) (prepare) is to be shown for the dosing
device connected to the MSB.
At program start
on | off (Default value: on)
If this option is switched on then at each program start the request to pre-
pare the dosing device will appear.
On attaching an exchange/dosing unit
on | off (Default value: on)
If this option is switched on then at each attaching of an exchange or dos-
ing unit the request to prepare the dosing device will appear.