7 properties - lift positions (774) – Metrohm tiamo 2.3 Manual User Manual
Page 1473
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3
■■■■■■■■
1457
Number of positions
Shows the number of positions on the rack. If no rack is in position then
the display will be empty.
The parameters for the attached rack are defined on the following 2 tabs:
■
Lift positions
Defines the work, rinse, shift and special position for Tower 1.
■
Special beakers
Display of the special beaker for the rack in position.
6.12.3.7
Properties - Lift positions (774)
Dialog window: Configuration
▶ Rack data ▶ Properties
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions
except those that are defined as (see Chapter 6.12.3.8, page 1458).
Work position
Work position for Lift 1. At this lift position the electrodes, stirrer and
buret tips are optimally positioned for work.
Range
0 ... 100 mm
Rinse position
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.
Range
0 ... 100 mm
Shift position
Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be
selected so that a safe movement across the entire rack is possible at any
time.
Range
0 ... 100 mm
Special position
Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solu-
tion.
Range
0 ... 100 mm