Metrohm tiamo 2.3 Manual User Manual
Page 1527
■■■■■■■■■■■■■■■■■■■■■■
7 Instruments
tiamo 2.3
■■■■■■■■
1511
defined in Rack table(see Chapter 6.12.2.1, page 1449) for general
Sample positions(see Chapter 6.12.3.2, page 1452) and Special beak-
ers(see Chapter 6.12.3.4, page 1454) will be compared with the Min.
beaker radius. If this minimum beaker radius is undercut then the run
will be stopped and an error message will appear. With off no check will
be made.
Selection
1.0 … 100.0 mm | off
Default value
off
Lift rate
Lift speed for manual operation.
Range
3 ... 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.
Range
100.0 ... 300.0 mm
814 USB Sample Processor XL
Selection
166.0 mm
Default value
166.0 mm
815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864
Balance Sample Processor
Selection
196.0 mm
Default value
196.0 mm
Swing Head
This parameter is only shown when a swing head is mounted on the
tower.
Serial number
Shows the serial number of the swing head.
Swing position
Lift position valid for all 4 external positions, at which the swing head
turns to the external positions.
Range
0 ... 235 mm
Default value
0 mm
Rinse position
Rinse position valid for all 4 external positions.