5 Method
tiamo 2.3
■■■■■■■■
797
drift correction is the period between end of conditioning and end of
titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size
of the titration cell in order to prevent any overflow.
Titrando
Range
0.000000 ... 9999.99 mL
Default value
20.000 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando
Range
1 ... 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Range
0 ... 999 s
Default value
0 s