8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1596
■■■■■■■■
tiamo 2.3
Current position
Shows the current position of the robotic arm in °.
External position
Selection of the external position to which the current robotic arm posi-
tion is to be assigned.
Range
1 ... 4
Default value
1
[Assign]
Trigger assignment. During the assignment the cursor is displayed as hour-
glass.
8.4.4.5
Sample changer - Pumping
Program part: Manual control
Here the pumps connected to the tower can be switched on and off.
Pump 1
Manually control pump 1 or valve 1 on the selected tower.
Operation
Selection whether the pump or valve should be switched on/off manually
or whether they should be switched on for a defined duration and then
automatically be switched off.
Selection
On/Off | Duration
Default value
On/Off
Duration
Entry of the duration during which the pump/valve should remain
switched on. This field only appears for Operation = Duration.
Range
1 ... 999999 s
Default value
60 s
[Start]
Start pump 1. In the status display the running time since start is shown.
[Stop]
Stop pump 1.
Pump 2
Manually control pump 2 or valve 2 on the selected tower.