5 Method
tiamo 2.3
■■■■■■■■
627
Start measured value
Once the start measured value is reached, the dosing of the start volume
is stopped and the next start condition is processed or the titration is star-
ted. If the start measured value is achieved by the addition of a start vol-
ume, then titration will start directly.
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
Start slope
Note
Only displayed for Titrando and 855.
Start slope
When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.
Range
0 ... 999 mV/mL
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum