2 sample solutions (tc conductivity) table, 1 sample solutions (tc conductivity) table, Sample solutions (tc conductivity) table 2 – Metrohm tiamo 2.3 Manual User Manual
Page 1458
6.11 Subwindow Sample solutions (TC conductivity)
■■■■■■■■■■■■■■■■■■■■■■
1442
■■■■■■■■
tiamo 2.3
This data can either be edited manually or determined with the help of the
command MEAS TC Cond. A sample solution (TC conductivity) can be
selected in the command MEAS Cond, in order to compensate for tem-
perature deviations during the conductivity measurement.
6.11.2
Sample solutions (TC conductivity) table
6.11.2.1
Sample solutions (TC conductivity) table
Subwindow: Configuration
▶ Sample solutions (TC conductivity)
Contents
In the default settings, the following data is shown in the table for the cal-
culation of the temperature coefficients:
Name
Name of the sample solution whose temperature coefficient was deter-
mined for conductivity.
Start temperature
Temperature at the start of the measurement.
Stop temperature
Temperature at the time of the end of the measurement.
TC min.
Minimum temperature coefficient in the temperature interval of the Che-
byshev function, with reference to a temperature of 25 °C.
TC max.
Maximum temperature coefficient in the temperature interval of the Che-
byshev function, with reference to a temperature of 25 °C.
c0 ... c4
Coefficients of the Chebyshev polynomial for the calculation of the func-
tion.
Assignment date
Moment at which the data for the calculation of the temperature coeffi-
cient was entered or determined.
Assignment method
Name of the method with which the automatic value assignment took
place. If the value has been entered manually then manual will be
shown.