2 meas u - general/hardware, General/hardware – Metrohm tiamo 2.3 Manual User Manual
Page 963
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
947
Designation
Description
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measur-
ing point list is reached
.MA.MEA
Maximum measured value (U
ind
) in mV
.MA.TEM
Temperature for the maximum measured value in °C
.MA.TIM
Time in s until the maximum measured value is reached
.MI.MEA
Minimum measured value (U
ind
) in mV
.MI.TEM
Temperature with minimum measured value in °C
.MI.TIM
Time for the minimum measured value in s
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.SLO
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
5.6.4.4.2
MEAS U - General/Hardware
Tab: Method
▶ MEAS U ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are
defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.