4 properties - rack, Properties - rack 8 – Metrohm tiamo 2.3 Manual User Manual
Page 1554
7.13 864 Balance Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
1538
■■■■■■■■
tiamo 2.3
[Edit]
Opens the dialog window External position # (see Chapter 7.11.8,
page 1525) in which the parameters for the external position selected
from the table can be edited.
7.13.4
Properties - Rack
Tab: Configuration
▶ Device ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Rack
On the tab Rack the rack-specific data of the attached rack are shown.
Rack name
Shows the name of the rack attached. If no rack is in position then "-----"
is shown.
Rack code
Shows the rack code of the rack attached. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the
Sample Processor in order to recognize the rack. If no rack is in position
then the display will be empty.
Number of positions
Shows the number of positions on the rack. If no rack is in position then
the display will be empty.
Shift rate
Sample rack speed of rotation for manual operation.
Range
3 ... 20 °/s
Default value
20 °/s
[Rack Data]
Opens the dialog window Rack data (see Chapter 6.12.3.1, page 1451)
or Rack data (774) (see Chapter 6.12.3.6, page 1456) in which the data
of the rack attached can be displayed and edited.
[Initialize rack]
Initializes the attached rack. This resets the rack, the lift and the robotic
arm, reads out the rack code and transfers the corresponding rack data to
the Sample Processor.