6.2 Administration
■■■■■■■■■■■■■■■■■■■■■■
1304
■■■■■■■■
tiamo 2.3
Security settings can be exported and imported. This means that these set-
tings can be exchanged between different client/server systems.
With the menu item File
▶ Print (PDF)... ▶ Security settings, the
security settings can be put out as PDF file.
Settings according to 21 CFR Part 11
Settings according to 21 CFR Part 11
on | off (Default value: off)
If this check box is activated, then the default settings pursuant to 21 CFR
Part 11 will be activated on all tabs. At the same time, those check boxes
which are required to be switched on pursuant to 21 CFR Part 11 will be
displayed against a gray background, i.e. they can no longer be edited.
Tabs
The security settings are configured on the following tabs:
■
Login/Password protection
Settings for login and password protection.
■
Audit Trail/Changes
Activation/Deactivation of the Audit Trail and comments when meth-
ods, determinations or sample data are modified.
■
Signatures
Options for electronic signatures
■
Default reasons
Definition of reasons for signing and modifying methods, determina-
tions and sample data.
6.2.2.2
Login/Password protection
Tab: Configuration
▶ Tools ▶ Security settings ▶ Login/Password protection
Definition is made on the tab Login/Password protection as to
whether the user must log in with name only or with name and password
and how the password must be monitored and constructed.
Login policies
Enforce login with user name
If this check box is activated then every time that the program starts, the
login window will appear in which the user must enter his/her user name.
If this check box is deactivated no login is necessary and the user logged
in under Windows will be adopted as user name. In this case all subse-
quent parameters are deactivated.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)