5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1054
■■■■■■■■
tiamo 2.3
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
Smoothing
Number of neighboring pixels with which the light intensity is determined
for each measurement pixel.
Range
0 ... 100 Pixels
Default value
0 Pixels
Tab for the 089 Photometer
The parameters for the measuring procedure can be defined on this tab.
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is selec-
ted. The measurement is canceled as soon as the defined Signal drift
or the Stop measured value is reached or a Max. waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled.
Measurement without drift control
The measurement is carried out without drift control if this option is
selected. The measurement continues until one of the stop criterions
Measuring time or Stop measured value is met. The correspond-
ing parameters are not visible if this option is disabled.
Measurement with drift control
Signal drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.