5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
676
■■■■■■■■
tiamo 2.3
Sensor
Selection of a sensor of the Conductivity sensor type from the sensors
configured in the sensor table. The calibration data for the sensor will be
adopted for the determination.
Selection
Sensor name | Conductivity sensor
Default value
Conductivity sensor
Temperature measurement
Type of temperature measurement.
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.