Metrohm tiamo 2.3 Manual User Manual
Page 867
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
851
Conditioning
on | off (Default value: on)
Enable/Disable the conditioning. If the conditioning is disabled, the follow-
ing parameters are not visible.
Start drift
The determination can only be started when the currently measured drift
is less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series tab will then switch to COND READY.
Range
1 ... 999 µg/min
Default value
20 µg/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction is the period between end of conditioning and end of
titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 999.9 µg/min
Default value
0.0 µg/min
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Range
1 ... 999999 s
Selection
off
Default value
off