6 tet - thermometric evaluation, 7 tet - additional evaluations – Metrohm tiamo 2.3 Manual User Manual
Page 935
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
919
5.6.3.9.6
TET - Thermometric evaluation
Tab: Method
▶ TET ▶ Properties... ▶ Thermometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the thermometric evaluation of the titration curve with
automatic endpoint recognition.
Evaluation start
Possibility to postpone the start of the evaluation. The evaluation starts
not before having added the solution volume.
Range
0.0 ... 50.0 mL
Default value
0.1 mL
Endpoints
Selection of the sorting for the display of the endpoints in the table.
Selection
Sort by volume (ascending) | Sort by peak
height (descending)
Default value
Sort by volume (ascending)
Table of endpoints
A maximum of 9 endpoints can be defined. The parameters for determin-
ing these endpoints are displayed in the window table which cannot be
edited and which contains the columns Reaction type and EP criterion.
[New]
Open the dialog window Endpoint # in which the parameters for a new
endpoint can be entered (see Chapter 5.6.3.9.9, page 922).
[Properties]
Open the dialog window Endpoint # in which the parameters for the
selected endpoint can be edited (see Chapter 5.6.3.9.9, page 922).
[Delete]
Delete the endpoint selected in the table.
5.6.3.9.7
TET - Additional evaluations
Tab: Method
▶ TET ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters