5 Method
tiamo 2.3
■■■■■■■■
811
Pause 2
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.5.3.4
KFT Upol - Control parameters
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
Endpoint
EP with
Measured value for the endpoint.
Range
-200.0 ... 200.0 µA
Default value
25.0 µA
Selection
off
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.