5 dosing - emptying, 6 dosing - dosing a fixed volume, Emptying – Metrohm tiamo 2.3 Manual User Manual
Page 1598: Fixed volume
8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1582
■■■■■■■■
tiamo 2.3
8.4.1.5
Dosing - Emptying
Program part: Manual control
Here the emptying of dosing units can be started and stopped. This tab
appears when selecting a single dosing device as well as with all dosing
devices with dosing units.
Note
This tab is not displayed for external dosing devices on the Titrinos
736, 751, 758, 799.
[Start]
Start emptying for the selected dosing device(s). The parameters defined
for the dosing units will be used.
[Stop]
Stop emptying for the selected dosing device(s).
8.4.1.6
Dosing - Dosing a fixed volume
Program part: Manual control
Here the dosing of a preset volume can be started and stopped. This tab
only appears if a single dosing device is selected.
Volume
Fixed volume that is to be dosed.
Range
0.100 ... 99999.9 mL
Default value
0.100 mL
Dosing rate
The volume is dosed at this rate. The maximum dosing rate depends on
the cylinder volume of the exchange unit or dosing unit used. If the
entered dosing rate is too high for the selected dosing device, it will auto-
matically be reduced during dosing to the largest possible value.
Note
The dosing rate should be reduced for viscous liquids.
Selection
maximum
Default value
maximum
Titrando, 814, 815, 855
Range
0.01 ... 166.00 mL/min