7 Instruments
tiamo 2.3
■■■■■■■■
1575
Data storage under legal control
Indicates whether the balance has its own data storage under legal con-
trol.
Only visible and editable with Sartorius balances
Selection
on | off
Default value
off
Remarks
Remarks about the device.
Input
1000 characters
Default value
empty
7.19.4
Properties - Settings
Tab: Configuration
▶ Device ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Settings
Barcode reader ID
Shows the identification of the barcode reader.
Input target
Selection of the field into which the output from the barcode reader is to
be entered.
Selection
Active input field | Method | ID1…16 | Sample
size | Sample size unit | Sample position
Default value
Active input field
[Connect]
Establishes a connection to the barcode reader. The dialog window
Establish connection (see Chapter 7.19.6, page 1577) appears.
[Disconnect]
Disconnects the connection to the barcode reader (only then the device
can be deleted out of the configuration).
7.19.5
Properties - GLP
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in
the dialog window Select date (see Chapter 2.5.1, page 89).