5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1018
■■■■■■■■
tiamo 2.3
Temperature
Measuring temperature
Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.
712
Range
-170.0 ... 500.0 °C
Default value
20.0
856
Range
-20.0 ... 150.0 °C
Default value
20.0
Reference temperature
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).
712
Range
-170.0 ... 500.0 °C
Default value
20.0
856
Range
-20.0 ... 150.0 °C
Default value
20.0
Selection
off
off
The conductivity at the measuring temperature is displayed.
Temperature compensation
Definition of the type how the temperature compensation is to be carried
out.
Selection
Temperature coefficient | Sample solution
Default value
Temperature coefficient
Temperature coefficient
Manual entry of a constant temperature coefficient.
Sample solution
Selection of a function for the temperature coefficient from the table
Sample solutions (TC conductivity).
Temperature coefficient
The temperature compensation is calculated on the basis of the value
entered (constant).