6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
1330
■■■■■■■■
tiamo 2.3
Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.
Note
We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line conditions.
Pulse length
Duration of the transmitted pulse.
Range
100 ... 1000 ms
Default value
200 ms
Examples for 765 Dosimat
Parameters
Binary pattern
Function
Start Dos1
*******p******
Starts Dosimat 1 (Titrino via "acti-
vate")
Start Dos2
*****p********
Starts Dosimat 2 (Titrino via "acti-
vate")
Start Dos*
*****p*p******
Starts Dosimat 1 and 2 (Titrino via
"activate")
6.3.3.4
Templates - Conductivity standards
6.3.3.4.1
Manage conductivity standards
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Conductivity stand-
ards
▶ Conductivity standards
Conductivity standards can be defined in the dialog window, which
can be selected with the command CAL Cond. The table cannot be edi-
ted.
[New]
Creation of a new conductivity standard. The window Edit conductivity
standard opens, in which a new conductivity standard can be defined
(see Chapter 6.3.3.4.2, page 1331).
[Properties]
Edits the selected conductivity standard. The window Edit conductivity
standard opens, in which the conductivity standard can be edited (see
Chapter 6.3.3.4.2, page 1331).