5 meas ipol - additional measured values, Additional measured values – Metrohm tiamo 2.3 Manual User Manual
Page 985
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
969
Value for the upper limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Window = Time
Range
1 ... 999999 s
Default value
999999 s
5.6.4.5.5
MEAS Ipol - Additional measured values
Tab: Method
▶ MEAS Ipol ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.
[New]
Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 927).
[Properties]
Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 927).
[Delete]
Delete the calculated measured value selected in the table.