8 stat u - evaluations, Evaluations – Metrohm tiamo 2.3 Manual User Manual
Page 923
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3
■■■■■■■■
907
5.6.3.8.7.8
STAT U - Evaluations
Tab: Method
▶ STAT U ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following methods for the evaluation of titration curves can be activa-
ted and defined on this tab:
Rate evaluation
Rate evaluation
on | off (Default value: off)
If this check box is activated, than a maximum of 9 time slots can be
defined in which the mean dosing rate can be determined by linear
regression. The defined time slots are shown in the window table and can
be edited with the following buttons:
[New]
Open the dialog window Evaluation window # in which the parameters
for a new time slot can be entered (see Chapter 5.6.3.8.3, page 857).
[Properties]
Open the dialog window Evaluation window # in which the parameters
for the new time slot can be edited (see Chapter 5.6.3.8.3, page 857).
[Delete]
Delete the time slot selected in the table.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.8.7.10, page 910).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.8.7.10, page 910).