5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1188
■■■■■■■■
tiamo 2.3
Titrando, 855
Range
0.0 ... 2000.0 mV
Default value
2.0 mV
Upper limit
Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
Upper hysteresis
The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.
Titrando, 855
Range
0.0 ... 2000.0 mV
Default value
2.0 mV
Action
Selection of the event to be triggered if the upper or lower limit is excee-
ded:
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Default value
none
Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is execu-
ted.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a
message appears. As soon as the monitored measured value is again
within the limits (including hysteresis) reagent dosing can be resumed
by pressing the [Continue] button in this message box.