5 det upol - stop conditions, Stop conditions – Metrohm tiamo 2.3 Manual User Manual
Page 566
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
550
■■■■■■■■
tiamo 2.3
otherwise it is easy to add too large a volume in the equivalence point
region. The value should not be less than 1/100 cylinder volume.
Titrando, 855
Range
0.1 ... 9999.9 µL
Selection
off
Default value
off
Dosing rate
Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Temperature
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
5.6.3.2.5.5
DET Upol - Stop conditions
Tab: Method
▶ DET Upol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters