Metrohm tiamo 2.3 Manual User Manual
Page 1248
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1232
■■■■■■■■
tiamo 2.3
Calculation overview
All defined results are displayed in a table in the properties window of the
command CALC.
Result parameters
The parameters for the individual results are configured on the following 3
tabs:
■
Definition
Entering the formula, assigning the unit and switching the statistics on/
off.
■
Monitoring
Settings for result monitoring.
■
Options
Saving the result as common variable or as titer.
Command variables
The following command variables are generated in the method run by the
command CALC and can be used in formulas under the designation
'Command name.Variable designation':
Designa-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
Result variables
The following result variables are defined in the CALC command and can
be used in formulas under the designation 'RS.ResultName.Variable
designation':
Designa-
tion
Description
.VAL
Result value, optional, i.e. 'RS.RS01' = 'RS.RS01.VAL'
.ASD
Absolute standard deviation for the result
.MAX
Maximum value of the result
.MIN
Minimum value of the result
.MNV
Mean value of the result