5 Method
tiamo 2.3
■■■■■■■■
1155
Dosing rate
The rate at which the volume is dosed. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing device used. If the
entered dosing rate is too high for the selected dosing device, it will auto-
matically be reduced during dosing to the largest possible value.
Titrando, 846, 814, 815, 855, 864, 874
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
730, 774, 778, 789
Range
0.01 ... 160.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The dosing rate should be reduced for viscous liquids.
Filling rate
The buret will be filled at this speed after the dosing. The maximum filling
rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing
device, it will automatically be reduced during dosing to the largest possi-
ble value.
Titrando, 814, 815, 855, 864, 874
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
730, 774, 778, 789
Range
0.01 ... 160.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min