6 data for specified electrode types – Metrohm tiamo 2.3 Manual User Manual
Page 1354
6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
1338
■■■■■■■■
tiamo 2.3
Min. slope
Minimum permitted slope of the pH electrode.
Range
-999.9 ... 999.9 %
Default value
95.0 %
Max. slope
Maximum permitted slope of the pH electrode.
Range
-999.9 ... 999.9 %
Default value
103.0 %
Max. response time
Maximum permitted response time of the pH electrode. The voltage
measured after three minutes in stirred solution is used as a comparison
value for determining the response time. The response time is the time
after which the measured voltage has reached this comparison value
except for ±1 mV.
Range
0 ... 999 s
Default value
60 s
Min. offset potential
Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.
Range
-999 ... 999 mV
Default value
-15 mV
Max. offset potential
Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.
Range
-999 ... 999 mV
Default value
15 mV
6.3.3.6.6
Data for specified electrode types
Tab: Configuration
▶ Extras ▶ Templates ▶ Electrode types... ▶ [Proper-
ties]
▶ Very good electrode / Good electrode / Usable electrode
The three predefined electrode types Standard, Gel and Non-aqueous
are always present in the table and be neither edited nor deleted. The fol-
lowing limit values are defined for the evaluation of the electrodes of
these electrode types: