3 vx / r-33 conversion tube, 1 vx/r-33 conversion tube installation – INFICON HAPSITE ER Chemical Identification System User Manual
Page 73

2 - 35
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
2.8.3 VX / R-33 Conversion Tube
This procedure describes the steps required to prepare the HAPSITE to sample for 
VX or R-33 using the Conversion Tube. To detect VX or R-33, you must insert the 
Conversion tube following the guidelines below. 
The process of detecting VX or R-33 on the HAPSITE requires the conversion of 
VX or R-33 (high boiling point chemicals) to the G analog. The VX or R-33 molecule 
is broken at the sulfur bond when it comes in contact with a silver fluoride pad. The 
result is the formation of a volatile chemical ethyl methylphosphonofluoridate in the 
case of VX, or isobutyl methylphosphonofluoridate in the case of R-33. These 
compounds are detected by the HAPSITE as VX-G or R-33-G.
NOTE: Theoretically, other G agents can be detected with the VX conversion pad
in place. However, if other G agents are suspected, the sample should be 
also be run without the conversion tube in place. 
WARNING
Sulfur mustard cannot be detected with the conversion 
tube in place.
2.8.3.1 VX/R-33 Conversion Tube Installation
1
section 2.8.1, Probe Nut Assembly, on page 2-33
to ensure proper
ferrule orientation in the probe nut.
WARNING
Correct ferrule placement is critical to ensure a leak-free 
fit around the conversion tube.
2
Loosen the nut on the end of the probe approximately 1/4 to 1/2 of a turn. See
Figure 2-47
.
Figure 2-47 Loosening the Nut
