2 installing the headspace sampling system – INFICON HAPSITE ER Chemical Identification System User Manual
Page 51
2 - 13
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Transfer Line Insulation (INFICON part number 931-405-P1) — This foam sleeve
is used as insulation for the Transfer Line. The insulation extends battery life by
reducing the energy required to heat the line. Once heated, it helps to maintain the
temperature.
Replacement Needle Kit (INFICON part number 931-402-P1) — A new needle
assembly for when the original needle is worn, plugged, or broken.
HSS Carrying Shoulder Strap (INFICON part number 036-015) — A strap which
connects to the mounts on the sides of the HSS. It facilitates carrying the
instrument between the office and the field.
Y-Cable (INFICON part number 600-1131-P30) — A power cable to split the power
from the AC to DC power converter to both the HAPSITE and the HSS.
2.5.2 Installing the Headspace Sampling System
Required Parts
The following items are needed to install the HSS:
Headspace instrument,
Y-cable power splitter,
Transfer Line,
Carrier Gas,
Charged Battery.
Procedure
1
Install a fully charged battery into the HAPSITE.
2
Unplug the AC to DC power converter power supply from the HAPSITE. See
Figure 2-10
.
Figure 2-10 Power Supply to HAPSITE from AC To DC Power Converter
3
Connect the single end of the Y-cable (see
Figure 2-11
) to the power supply
connector removed in step 2.