5 instrument overview, 6 description of subsystems, 1 gas chromatograph – INFICON HAPSITE ER Chemical Identification System User Manual
Page 32: For more
1 - 4
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
1.5 Instrument Overview
The HAPSITE system is comprised of two modules:
the HAPSITE, also known as the Analytical Module,
the Service Module.
The HAPSITE is comprised of many systems and subsystems.
Figure 1-1
shows
a diagram of the major subsystems. These subsystems include the several pumps
used to provide flow and vacuum.
The Service Module components are identified in
Figure 1-1
, everything else is
housed in the HAPSITE. The Service Module and the HAPSITE module contain a
Vacuum Interconnect Valve and electrical connectors through which their vacuum
systems join. The modules communicate when the two modules are coupled
together.
Figure 1-1 Major HAPSITE Subsystems
1.6 Description Of Subsystems
The HAPSITE is comprised of the following subsystems:
Gas Chromatograph, see
Mass Spectrometer, see
.
Vacuum System, see
.
Electronic Systems, see
.
1.6.1 Gas Chromatograph
The HAPSITE’s GC system utilizes nitrogen as the carrier gas to transport analytes
through a column, a narrow-bore fused silica tube 15 meters in length, and then on
to the detector. The nitrogen is referred to as the mobile phase.