At the end of the Headspace run, a prompt will appear. When prompted Insert
needle into a clean vial and press Continue to Purge
, place the needle into
the empty vial. When finished, touch the Continue button or using the arrow
keys
, highlight Continue. Push OK SEL.

Figure 3-118 Purging

NOTE: If purge is skipped, a full Headspace run must be completed to flush

the system. Once purge has been skipped, it is not possible to
re-access it.

15

The METHOD FINISHED message will appear when the purge method has
ended. See

Figure 3-119

.

Figure 3-119 Sample Method Finished for HSS

NOTE: Another Analyze (GC/MS) method can be started immediately after one

has completed. Depending on the temperature profile, the column may
need to cool before another run will begin.

NOTE: Refer to

Section 3.5.1, View Results/View Reports, on page 3-32

for more

information on reviewing the data.