4 display function – INFICON HAPSITE ER Chemical Identification System User Manual
Page 405
12 - 9
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
12.4 Display Function
The Display button in the Data Files section of both Calibrate and ID Unknowns
shows the selected Date File’s Spectrum. This feature is beneficial when reviewing
and revising identifications, selecting spectral peaks, adding to a library and
manually integrating peak areas. See
Figure 12-7 Calibration Display File
To display a chromatogram while in Calibrate or ID Unknowns, select the file by
highlighting with the Left Mouse Button and selecting the Display button.
The controls of the Display function are similar to the data review functions for
working with RICs (Reconstructed Ion Chromatograms) and Mass Spectra. Data
Review is described in
. Additional features are described
below.
With the Build/Edit Template option selected in the Calibrate window, the
Display button will show the peaks detected and selected for the library template
when a peak search is performed. See
. Peaks selected for the
template are indicated on the plot with a T. The quant mass for each analyte is
displayed on the RIC plot. The T symbol will be the same color as the mass number
color displayed in the lower right corner of the RIC graph. If a peak has not been
detected, the mass Spectrum of that peak can be manually selected and the
compound it represents added. After selecting the Spectrum for the compound to