INFICON HAPSITE ER Chemical Identification System User Manual
Page 355

11 - 17
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
HINT: When editing from the keyboard, the Alt+Down Arrow key combination will
"drop down" the menu in the Inlet State column when this option is
highlighted.
The following parameters are found on the Inlet States page:
Start . . . . . . . . . . . . . . . . . . . . . . . . . This is the start time of each Inlet State and is
not programmable. The start of the Run Time
actually is the start of the first Injection event.
This being the case, there can be several
events occurring before the actual start of the
Run Time. Each Inlet State starts at the end
of the duration of the previous state.
Duration. . . . . . . . . . . . . . . . . . . . . . Displays the run time, in minutes and
seconds, of each Inlet State event.
Volume. . . . . . . . . . . . . . . . . . . . . . . Displays the volume of sample that will be
collected for the Concfill or Loopfill event.
Inlet State . . . . . . . . . . . . . . . . . . . . The order of the entire Method is
programmed via the Inlet States page and
specifically via the Inlet State column. A set
of pre-programmed Inlet States exist and are
selected from the drop down menu in the
Inlet State column. Highlighting an Inlet State
will show the Valve States used for that Event
(see
Figure 11-15
). An Inlet State of Other
can be selected, which allows for a custom
set of Valve States to be set for that particular
Event (see
Figure 11-16
). The selection of
some Inlet States is dependent on the
configuration or placement of this Event
within the Method. Some Valve States cannot
be changed for certain Inlet States.
NOTE: The list of Valve States may also include pumps, heaters and fans.
Selecting a component on the Valve States list tells the system to
activate the component. When the box is not checked, it tells the
system to deactivate the valve or component.